Production process analysis: a review of the device history record confirmed that the device was manufactured and tested according to relevant procedures, and shipped according to manufacturer's specifications.User (surgeon and patient) information analysis: on (b)(6) 2024 apifix was notified that patient (b)(6) (pas (b)(6) who was a patient in dayton is planned to be converted to fusion in another facility (north carolina) by another physician.On (b)(6) 2024 additional information was received.According to the reporter, "the patient's kyphosis has gotten so bad.In addition, the patient's curves are still progressing to about 50°.It is unsure at this time when the patient is having the fusion." reoperation events are a known risk that was assessed and recorded by the product risk management file.The risk of curve progression has been assessed and found to be acceptable.The current rate for all relevant criteria (curve progression, insufficient curve correction, imbalance) is in line with the rate reported in the literature for this type of complication as described in the company's cer (clinical evaluation report).The risk has been quantified, characterized, and documented as acceptable within a full risk assessment the explanted device is expected to be returned to the manufacturer and will be evaluated; upon its completion, once new information comes to light, then a supplemental medwatch report will be submitted.
|
|
Device was returned to manufacturer for analysis.Engineering analysis: the explanted device was returned to manufacturer, and was subjected to cleaning, steam sterilizing, and then went under engineering evaluation.The device was fractured in the pole component around the mid-point of the pole.Post-fracture, the device remained implanted as the fractured side of the pole showed signs of wear.The fracture plane was worn not in a completely smooth form, in the center of the fracture the plane was shiny and smooth, and the edges were uneven with ridges thus fracture mode is not able to be identified.The extender was removed, and the spherical rings were inspected for wear.Scratches were observed on the spherical ring and spherical ring housing of the pole.No wear was visibly observed on the spherical ring of the base.During the inspection, inside the apifix screw is the tip of lp nut leader tool, apparently during the removal, the tool broke.Curve progression can result from several reasons, such as device misplacement, operating on patients out of the approved indication, implant size selection, extender misalignment, screw misplacement /migration, screw pull-out, implant breakage, infection, ratchet malfunction, and in some cases for no apparent reason.The progression can be either or both of the primary or secondary curve and may also be reported together with pain and or spine imbalance.Cause for the secondary curve progression could not be determined.
|