| Model Number |
127-728/26 |
| Medical Device Problem Code |
Unclear Information (4052)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A field safety notice and capa was initiated as corrective action.
|
| |
|
Event or Problem Description
|
|
"customer performed surgery according to surgical technique.Trial reduction was done with 126 xl neck segment, which was then selected for definitive implant (r4 w/ xl neck).When circulating nurse showed the implant label to surgeon before opening the package, the size marking on the label ("x-large r4") was misinterpreted as "xl-neck r4".Stem was then implanted and it was noticed only after the cement was cured that it had standard neck but not xl." [customer].
|
| |
|
Event or Problem Description
|
|
"customer performed surgery according to surgical technique.Trial reduction was done with 126 xl neck segment, which was then selected for definitive implant (r4 w/ xl neck).When circulating nurse showed the implant label to surgeon before opening the package, the size marking on the label ("x-large r4") was misinterpreted as "xl-neck r4".Stem was then implanted and it was noticed only after the cement was cured that it had standard neck but not xl." [customer].
|
| |
|
Additional Manufacturer Narrative
|
|
A field safety notice and capa was initiated as corrective action.This is the final supplemental report, the complaint is closed.
|
| |
|