MAUDE Adverse Event Report: BIOMET 3I T3 PRO TAPERED IMPLANT 6/5MM (D) X 11.5MM (L); DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I T3 PRO TAPERED IMPLANT 6/5MM (D) X 11.5MM (L); DENTAL IMPLANT Back to Search Results
Catalog Number T3PT6511
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Hypersensitivity/Allergic reaction (1907)
Date of Event 01/04/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient is allergic to titanium, therefore the implant was removed.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A4: weight unknown / not provided e1: last/given name unknown / not provided product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 67.H10: narrative/data was updated.Zimvie received the reported implant for evaluation.Visual evaluation of the as returned device identified the implant with signs of use.No apparent malfunction was identified.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number 2022030668.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2022030668 for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are customer error and patient factors (material selection.Therefore, based on the available information, device malfunction did not occur.The reported event was non-verifiable with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T3 PRO TAPERED IMPLANT 6/5MM (D) X 11.5MM (L)
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key18639955
Report Number0001038806-2024-00180
Device Sequence Number7973778
Product Code DZE
UDI-Device Identifier00844868046820
UDI-Public(01)00844868046820(17)270419(10)2022030668
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K213672
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberT3PT6511
Device Lot Number2022030668
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/15/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/05/2024
Supplement Report FDA Received Date07/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-