MAUDE Adverse Event Report: SYNTHES GMBH 2.8MM THRDD GUIDE F/ 3.5MM SCR VA AND LCP; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.8MM THRDD GUIDE F/ 3.5MM SCR VA AND LCP; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Catalog Number 03.133.004
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2024 during an unknown procedure, the threaded drill guides in question would not thread into the proximal humerus locking plate.The surgeon tried with multiple threaded drill guides and all had trouble threading into the plate.The procedure was completed successfully without any surgical delay.No adverse patient impact was reported.No further information is available to report.This report is for a 2.8mm thrdd guide f/ 3.5mm scr va and lcp.This is report 6 of 10 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: initial reporter is a synthes employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D10: therapy date for all concomitant products is (b)(6) 2024.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.8MM THRDD GUIDE F/ 3.5MM SCR VA AND LCP
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18640019
Report Number8030965-2024-01922
Device Sequence Number8042035
Product Code FZX
UDI-Device Identifier10886982269973
UDI-Public(01)10886982269973
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.133.004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/22/2024
Supplement Date Received by Manufacturer03/12/2024
Initial Report FDA Received Date02/05/2024
Supplement Report FDA Received Date03/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2.8 THRDD GUIDE F/ 3.5 SCR VA AND LCP; 2.8 THRDD GUIDE F/ 3.5 SCR VA AND LCP; 2.8 THRDD GUIDE F/ 3.5 SCR VA AND LCP; 2.8 THRDD GUIDE F/ 3.5 SCR VA AND LCP; 2.8 THRDD GUIDE F/ 3.5 SCR VA AND LCP; 2.8 THRDD GUIDE F/ 3.5 SCR VA AND LCP; 2.8 THRDD GUIDE F/ 3.5 SCR VA AND LCP; 2.8 THRDD GUIDE F/ 3.5 SCR VA AND LCP; 2.8 THRDD GUIDE F/ 3.5 SCR VA AND LCP
Patient SexUnknown
-
-