• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 EMSYS SHL 3HOLE 54; HIP IMPLANT : ACETABULAR CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 EMSYS SHL 3HOLE 54; HIP IMPLANT : ACETABULAR CUP Back to Search Results
Catalog Number 471054300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377); Unspecified Musculoskeletal problem (4535)
Date of Event 01/03/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Subject id: (b)(6); study no: dots.Clinical notification received for revision due to periprosthetic fracture.Date of implant: (b)(6) 2023; date of revision: (b)(6) 2024; (left hip).Treatment: stem and head were revised.Medical records review (22_jan_2024_08_09_04_18105): on (b)(6) 2024, patients left total hip was revised to address chronic infection with progression of periprosthetic osteolysis of both femoral and acetabular bone.During the revision surgical procedure, it was identified that the femoral stem was loose, that it had only a fibrous ingrowth.There was an osteolytic defect beneath the superior aspect of the acetabular cup, but the cup was well-fixed.There was osteolysis of the femoral canal.The stem, femoral head, cup, liner and two acetabular screws were all explanted.A temporary antibiotic spacer, made up of an antibiotic cemented depuy summit stem, was cemented in place along with stimulant pellets, and paired with a bipolar head.Two synthes cerclage cables were placed on the proximal femur to address a calcar fracture that occurred intraoperatively during broaching for the spacer summit stem.The patient tolerated the procedure well, with no other complications apart from the calcar bone fracture caused by the summit broach.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMSYS SHL 3HOLE 54
Common Device Name
HIP IMPLANT : ACETABULAR CUP
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18641169
Report Number1818910-2024-02328
Device Sequence Number8240383
Product Code LPH
UDI-Device Identifier10603295500612
UDI-Public10603295500612
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K221636
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number471054300
Device Lot Number4290151
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2024
Supplement Date Received by Manufacturer02/08/2024
Initial Report FDA Received Date02/05/2024
Supplement Report FDA Received Date02/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE HD CER 40OD +5; EMSYS LNR AOX +4N 54X40; PINN CAN BONE SCREW 6.5MMX20MM; PINN CAN BONE SCREW 6.5MMX35MM; SUMMIT POR TAPER SZ5 HI OFF
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexMale
Patient Weight118 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-