Catalog Number 447202
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd kiestra inoqula there was false negative.There was no report of impact to patient or user.
 
Additional Manufacturer Narrative
H6: investigation summary: this statement is to summarize the investigation of a complaint involving the bd kiestra inoqula (material # 447202 - serial #(b)(6)).According to information provided, ino samples were not inoculated.No other issues were reported.During the investigation, the field service engineer (fse) observed the issue and cleaned the pdv sensor and calibrated the pipette on the pdv using the teach button, which changed the value from 172 to 158.Later the fse verified all values concerning the zeus 2, template editor, and auto tech are verified (158).Also verified the pipettor alignment and functionality.The pdv sensor was tested by catching drops from the pipettor multiple times, resulting in an error each time; the samples were not inoculated, and the dishes were directed to the error stacker.The lid lifter vacuum cup was tested by drilling holes in the lids, which resulted in a vacuum error with each attempt.Following this the fse installed a camera to observe the pdv sensor and dispensing activities in future tests.Approximately 40 samples were tested, with three dishes per sample, and footage from the camera was reviewed.The fse confirmed that the pipettes are dispensing correctly, and the tip is not causing interference with the pdv sensor.Following this action, no further issues were encountered, and the instrument is working properly as per specifications.This is a confirmed complaint of the bd product.Review found no new trends, risks, or hazards were identified because of this complaint.The issue in this complaint does not require the initiation of a corrective and preventative action (capa).A design history record (dhr) review is not required for this complaint.The complaint was evaluated via other elements of the investigation.Bd quality will continue to closely monitor for trends associated with this issue.
 
Event or Problem Description
It was reported when using the bd kiestra inoqula there was false negative.There was no report of impact to patient or user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD KIESTRA INOQULA
Common Device Name
DEVICE, MICROTITER DILUTING/DISPENSING
Manufacturer (Section D)
BD KIESTRA LAB AUTOMATION
6 marconilaan
drachten KY
Manufacturer (Section G)
BD KIESTRA LAB AUTOMATION
6 marconilaan
drachten KY
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18641224
Report Number3010141591-2024-00001
Device Sequence Number20886169
Product Code JTC
UDI-Device Identifier00382904472023
UDI-Public(01)00382904472023
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number447202
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2024
Supplement Date Received by Manufacturer05/08/2024
Initial Report FDA Received Date02/05/2024
Supplement Report FDA Received Date05/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-