• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number TECT1612AR
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Adhesion(s) (1695); Cellulitis (1768); Purulent Discharge (1812); Erythema (1840); Fistula (1862); Hemorrhage/Blood Loss/Bleeding (1888); Unspecified Infection (1930); Inflammation (1932); Pain (1994); Perforation (2001); Seroma (2069); Hernia (2240); Peritonitis (2252); Discomfort (2330); Impaired Healing (2378); Unspecified Tissue Injury (4559); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient's attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an inguinal hernia.It was reported that after implant, the patient experienced pain, infection, recurrence, inflammation, adhesions, bleeding, fistula, seroma, perforation.Post-operative patient treatment included hernia repair with new mesh, medication, lysis of adhesions, omentectomy, exploratory laparotomy, placement of vascular graft, primary hernia repair, abdominoplasty, and excision of infected mesh.
 
Additional Manufacturer Narrative
Additional info: b2, b5, b7, d10 (product id - abstack30, lot # - n1k0832x), h6 (patient codes, ime e2402: skin sinuses), <(>&<)> additional codes medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
The patient's attorney alleged a deficiency against the device.The product was used for a laparoscopic therapeutic treatment of an inguinal hernia.It was reported that after implant, the patient experienced pain, infection, recurrence, inflammation, adhesions, bleeding, fistula, seroma, perforation, dense obliterative peritonitis, open wound with erythema, swelling, tenderness, purulent sanguineous drainage, skin sinuses, <(>&<)> cellulitis.Post-operative patient treatment included hernia repair with new mesh, medication, lysis of adhesions, omentectomy, exploratory laparotomy, placement of vascular graft, primary hernia repair, abdominoplasty, excision of infected mesh, abdominal wall reconstruction, bilateral modified myofascial flap advancement, soft tissue flap advancement and rearrangement abdominoplasty, debridement of subcutaneous tissue/skin, hospitalization, antibiotics, iv antibiotics, component separation, <(>&<)> use of jp drains.
 
Additional Manufacturer Narrative
Correction: g1 & h6 (added patient code).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARIETEX
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112
8006464633
MDR Report Key18641829
Report Number9615742-2024-00146
Device Sequence Number17924334
Product Code FTL
UDI-Device Identifier10884521179967
UDI-Public10884521179967
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K982532
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/30/2016
Device Model NumberTECT1612AR
Device Catalogue NumberTECT1612AR
Device Lot NumberSLG00443
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2024
Supplement Date Received by Manufacturer06/18/2025
07/02/2025
Initial Report FDA Received Date02/05/2024
Supplement Report FDA Received Date06/24/2025
07/03/2025
Date Device Manufactured08/01/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 COMMENT
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight73 KG
-
-