| Catalog Number |
105-3600 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
Loss of Range of Motion (2032)
|
| Date of Event |
01/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision occurred approximately 6 years after the primary surgery, which occurred in (b)(6) 2018.The primary usage form was not able to be located, as the product was distributed through another vendor.No fall or other trauma reported.The cause for the revision was the glenosphere disassociating with the baseplate, seemingly over a long period of time.Surgeon notes indicate mobility and range of motion had been compromised.36 mm centered glenosphere and 36 mm + 6 standard humeral cup were explanted; the included part numbers are likely but not absolute as no primary usage form could be obtained.These parts were replaced with a 36 mm eccentric glenosphere, and 36 mm + 6 humeral stability cup.
|
| |
|
Additional Manufacturer Narrative
|
|
Revision occurred approximately 6 years after the primary implant.The glenosphere dissociated with the baseplate over a period of years.No actual, implied, or suspect link to fx product.
|
| |
|