• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5 12.6
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/05/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A4-a6: unk.Claim # (b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 12.6mm vticm5_12.6 implantable collamer lens of diopter -7.5/1.0/082 (sphere/cylinder/axis) into the patient's left eye (os) on (b)(6) 2023.On (b)(6) 2024 the lens was exchanged for a shorter length lens due to excessive vault.The problem was resolved.The cause of the event was reported as unknown.
 
Additional Manufacturer Narrative
H3 - device evaluation: the lens was returned with moisture in a microcentrifuge vial.Visual inspection haptic bent to the lens.Dimensional inspection found the lens to be within specifications.Claim #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key18645453
Report Number2023826-2024-00515
Device Sequence Number17920022
Product Code QCB
UDI-Device Identifier00840311307777
UDI-Public00840311307777
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model NumberVTICM5 12.6
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer08/15/2024
Initial Report FDA Received Date02/05/2024
Supplement Report FDA Received Date08/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45 - LOT# UNK; FOAM TIP PLUNGER - LOT# UNK; INJECTOR MODEL: MSI-PF - LOT# UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexUnknown
-
-