| Catalog Number |
MZ9283 |
| Medical Device Problem Codes |
Leak/Splash (1354); Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bd maxzero multi-fuse extension set with needleless connector luer lok broke the following information was received by the initial reporter with the verbatim: the leur lock on the 4 lumen y tubing frequently cracks and breaks putting patients at risk for infection or disconnection of critical medications.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported the luer lock is cracking and returned one used sample and one unopened sample, bot both of material mz9283 and lot 22119050.The sets were visually examined for defects and abnormalities.The used set luer collar was clearly broken in multiple pieces and was not attached to the set.The unopened set did not have any issues on visual examination.Upon testing and priming with water from a 10 ml bd syringe, no leaks or issues were found.The complaint of luer lock cracking is verified by the used set.The supplier for the luer was contacted about the issue.The root cause for the luer crack was not able to be found.Device history record review for model mz9283 lot number 22119050 was performed.The search showed that a total of 7,903 units in 1 lot number was built on 14nov2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|