MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
|
Back to Search Results |
|
| Model Number |
990063-020 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cryo ablation procedure, there was difficulty attempting to cannulate the right inferior pulmonary vein (ripv).Eventually the transeptal was lost.The mapping catheter was then removed and damage was noted; the mapping catheter was "stretched out." the mapping catheter was replaced to resolve the issue.Additionally, while attempting to advance the sheath, it would not advance into the groin.Another dilator was used without resolve.The sheath was removed from the wire and a bigger dissection was made into the tissue without resolve.The sheath tip was noted to be bent.The sheath was replaced to resolve the issue. the case was completed with cryo. no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the 990063-020 mapping catheter with lot number 227669017 was returned and analyzed.Damage was observed on the returned device.Due to the condition of the returned device, only visual inspection was performed.The loop shape was deformed and out of plane.The user may experience insertion difficulties due to this issue when introducing the mapping catheter into the balloon catheter.Visual inspection of the pebax tubing area showed the pebax tubing was intact with no apparent issues.No damage was observed along with the pebax tubing section of the mapping catheter.Visual inspection of the electrode(s) showed the electrode(s) was intact with no apparent issues.All electrodes exist on the loop section, and no cosmetic issue or anomalies were identified.Visual inspection of the shaft segment area showed the shaft was intact with no apparent issues.No kink or any other damage was observed along with the shaft of the achieve mapping catheter visual inspection of the introducer showed the introducer was intact with no apparent issues.No damage or any other issue was observed along with the introducer.Visual inspection of the lemo connector showed the lemo connector was intact with no apparent issues.No damage or any other issue was observed along with the lemo connector.In conclusion the reported issues was confirmed through testing.The mapping catheter failed the returned product inspection due to the out of plane and deformed loop.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cryo ablation procedure, there was difficulty attempting to cannulate the right inferior pulmonary vein (ripv).Eventually the transeptal was lost.The mapping catheter was then removed and damage was noted; the mapping catheter was "stretched out." the mapping catheter was replaced to resolve the issue.Additionally, while attempting to advance the sheath, it would not advance into the groin.Another dilator was used without resolve.The sheath was removed from the wire and a bigger dissection was made into the tissue without resolve.The sheath tip was noted to be bent.The sheath was replaced to resolve the issue. the case was completed with cryo. no patient complications have been reported as a result of this event.
|
| |
|