| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.6mm vticm5_12.6; -12.00/+3.0/081 (sphere/cylinder/axis) implantable collamer lens into the patient's right eye (od) on (b)(6) 2023.The patient experienced lens rotation.On (b)(6) 2023 the lens was exchanged with a same length but spherical lens.This resolved the problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Type of investigation code: 10- device evaluation: the lens was returned dry in a microcentrifuge vial with residue/debris on product.Visual inspection found haptic broken.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|