• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN LAPRO-CLIP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN LAPRO-CLIP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 8886848812
Medical Device Problem Codes Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant products: 8886848812, 8886848812 laproclip 6x12 x6 (lot#:n3c0685y) 8886848812, 8886848812 laproclip 6x12 x6 (lot#:n3c0685y) 8886848812, 8886848812 laproclip 6x12 x6 (lot#:n3c0685y) 8886848812, 8886848812 laproclip 6x12 x6 (lot#:n3c0685y) 8886848812, 8886848812 laproclip 6x12 x6 (lot#:n3c0685y).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic partial gastrectomy, prior to patient use, after opening the pack of six clip appliers, it was found that all six clips had been closed, or deformed and fell off, making it impossible to try them out.A new clip applier was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found that since the instrument was not received an rpa representative instrument c2l3101 was used for functional testing.The cartridge was loaded into the representative instrument and was applied to the test media with acceptable results.It was reported that there was a clip formation and the device component was disengaged.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when the cartridge is subjected to force or twisting during application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPRO-CLIP
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18647718
Report Number1219930-2024-00571
Device Sequence Number15515185
Product Code GCJ
UDI-Device Identifier10884521057654
UDI-Public10884521057654
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K925602
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8886848812
Device Catalogue Number8886848812
Device Lot NumberN3C0685Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2024
Supplement Date Received by Manufacturer05/08/2024
Initial Report FDA Received Date02/05/2024
Supplement Report FDA Received Date05/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-