| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a hair-like substance in the package.The event timing was outside of surgery.There is no harm or delay reported.Due diligence is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete g2: foreign country - japan multiple mdr reports were filed for this event, please see associated reports: 0001526350-2024-00194 0001526350-2024-00195 0001526350-2024-00196 0001526350-2024-00197.
|
| |
|
Additional Manufacturer Narrative
|
|
Multiple mdr reports were filed for this event, please see associated reports: 0001526350-2024-00194-1 0001526350-2024-00195-1 0001526350-2024-00196-1 0001526350-2024-00197-1 visual examination of the returned product/provided pictures identified a loose hair-like particular within the sterile packaging.The packaging did not meet the acceptance criteria per zpo 8.400 packaging materials inspection.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
There is no additional event information available.
|
| |
|