This device is classified as import for export, therefore 510k is not applicable.Model eg34-j10u-us is available in the usa with a 510k number k200090.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the shadow in image as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, our technician confirmed that the segment fluid damage, the ultrasound connector body cracked, the operation channel (primary) leak, the ultrasound connector body leak, the objective lens scratched, and the suction channel kink; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|