MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE ELECTROSURGICAL KNIFE KD-612

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE ELECTROSURGICAL KNIFE KD-612 Back to Search Results
Model Number KD-612L
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the tip of the electrosurgical knife fell off.The issue occurred during a therapeutic endoscopic submucosal dissection (esd) procedure.The fallen part was collected and the procedure was completed using a similar device.There was no reported patient harm or impact due to this event.
 
Additional Manufacturer Narrative
During device evaluation at olympus, it was found the cutting knife was broken.The tip of the knife was charred black and there were traces of electrical discharged.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause of the tip falling off could not be determined, likely factors causing the issue could have been the following: 1 - during tissue incision, discharges occurred due to the following factors.The energization time was long.The contact length between the tissue and the incision knife was short.2 - high heat was locally applied to the incision knife by the discharge, and the strength of the incision knife decreased (became thinner).3 - the incision knife with reduced strength was subjected to a load during tissue incision and broke.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the following ifu : when the electrosurgical unit is used in the coagulation mode, crack/detachment of the tip and the electrode or deformation/break of the cutting knife could occur, for example when the high-frequency output is set too high or the length of the contact between the cutting knife and tissue is too short.During treatment, always ensure that the slider slides on the handle smoothly and that the electrosurgical knife observed in the endoscopic image is normal.Should cracks or detachment of the tip or deformation/break of the cutting knife be detected during use, immediately shut off the power supply, discontinue the procedure, pull the slider and withdraw the endoscope from the patient with the cutting knife retreated in the insertion portion of this instrument.Do not continue using an abnormal electrosurgical knife to prevent perforation or hemorrhages.If the tip or cutting knife is detached be sure to collect it using grasping forceps.Aspirate fluids such as mucus that adhere to the electrode and/or the cutting knife, outer sheath and body cavity tissues.Patient injury such as punctures, hemorrhages, mucous membrane damage and thermal injury of tissue could result if output is activated when in contact with these adhering fluids.When current is discharged while the cutting knife is being separated from the mucosa under wet situation, it may break the cutting knife or crack the tip.Always operate the electrosurgical unit at the minimum output level and for the minimum time necessary to successfully complete the procedures.Excessive output level and time may result in patient injury, such as punctures, hemorrhages or mucous membrane damage.Application of high-voltage waveforms for extended periods increase the likelihood that it may break the cutting knife or crack the tip.When a high-voltage waveform has to be used, minimize the duration of current application.If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ELECTROSURGICAL KNIFE KD-612
Common Device Name
SINGLE USE ELECTROSURGICAL KNIFE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18648642
Report Number9614641-2024-00273
Device Sequence Number7187153
Product Code KNS
UDI-Device Identifier04953170453922
UDI-Public04953170453922
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K092309
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-612L
Device Lot Number17K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2022
Initial Report FDA Received Date02/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-