| Model Number |
30MM CENTRAL SCREW, MODULAR |
| Medical Device Problem Code |
Damaged Thread (4075)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 01/16/2024, it was reported by a sales representative via email that an ar-9560-24-2 arthrex univers revers modular glenoid system modulas baseplate 24 mm +2 lat and an ar-9561-30s univers revers modular glenoid system, central screw modular 30 mm cold-welded together on the press.When separating the modular glenoid system modulas baseplate from the screw, it stripped.This was discovered during a procedure on (b)(6) 2024.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint allegation was confirmed.One unpackaged ar-9561-30s was received for investigation.Functional testing was not performed due to the damaged to the device connector.The visual evaluation found that the hex was damaged, revealing nicks.The most likely cause for the reported failure is user error following the device's surgical technique.When two devices are intended to be threaded together, ensure that they are fully engaged prior to use.
|
| |
|