• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 24MM +2 LAT BASEPLATE, MODULAR; SHOULDR PROSTH, REVERSE CONFIG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 24MM +2 LAT BASEPLATE, MODULAR; SHOULDR PROSTH, REVERSE CONFIG Back to Search Results
Model Number 24MM +2 LAT BASEPLATE, MODULAR
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/15/2024
Type of Reportable Event Malfunction
Event or Problem Description
On 01/16/2024, it was reported by a sales representative via email that an ar-9560-24-2 arthrex univers revers modular glenoid system modulas baseplate 24 mm +2 lat and an ar-9561-30s univers revers modular glenoid system, central screw modular 30 mm cold-welded together on the press.When separating the modular glenoid system modulas baseplate from the screw, it stripped.This was discovered during a procedure on (b)(6) 2024.Additional information requested.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The complaint allegation was confirmed.One unpackaged ar-9560-24-2 serial/batch number (b)(6) was received for investigation.Functional testing was not performed due to damage to the returned mating part ar-9561-30s.The visual evaluation found that the baseplate connector hex was damaged, revealing nicks on the connector.The most likely cause for the reported failure is user error following the device's surgical technique.When two devices are intended to be threaded together, ensure that they are fully engaged prior to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
24MM +2 LAT BASEPLATE, MODULAR
Common Device Name
SHOULDR PROSTH, REVERSE CONFIG
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key18649159
Report Number1220246-2024-00780
Device Sequence Number12576635
Product Code PHX
UDI-Device Identifier00888867296497
UDI-Public00888867296497
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K173900
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24MM +2 LAT BASEPLATE, MODULAR
Device Catalogue NumberAR-9560-24-2
Device Lot Number15134150
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2024
Supplement Date Received by Manufacturer08/05/2024
Initial Report FDA Received Date02/06/2024
Supplement Report FDA Received Date08/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-