EXACTECH, INC. NV GXL LNR, +5LAT, 36MM G2-52/54MM CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
136-36-52 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Myalgia (2238); Osteolysis (2377); Implant Pain (4561)
|
| Date of Event |
01/16/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As part of the manufacturer's recall campaign, the patient presented himself for a check-up of the device implanted in 2013 left hip prosthesis.The x-ray control showed a clear decentering of the prosthetic head a very large osteolysis in the acetabulum as a sign of inlay wear.There is an additional finding pronounced lumbar spine syndrome in polymyalgia rheumatica.During the revision surgery on (b)(6) 2023, he was able to radiological findings are confirmed.Because the socket was loose and osteolysis occurred due to the size then the complete socket change to a revision hip cup with tab and iliac pin.As part of the replacement operation, the changing the prosthetic head.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending investigation.2491734 180-01-54 - crown cup,cluster-hole gr.54 h7: z-2128-2021.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported may have been the result of the orientation of the acetabular cup, edge loading/impingement, patient-related conditions, or any combination of these possibilities, which led to polyethylene wear and osteolysis after being implanted for 10+ years.It should also be noted that using both a lateralized liner and a group 2 acetabular liner with a 36mm femoral head creates a combination of the largest available head and the thinnest available acetabular liner (for group 2), which is a risk factor for early wear and lysis specified in the hhe.However, the reported wear and reason for revision cannot be confirmed as the devices were not available for evaluation and no pictures, radiographs, or medical records were provided at the time of this evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information, the patient involved meets the following risk criteria for early prosthesis wear and/or osteolysis as specified in the hhe: implanted with a lateralized liner, combination of largest available femoral head and/or thinnest available liner was used.The most likely underlying cause for the revision reported is a combination of risk factors specified in the hhe and being implanted for more than 10 years.However, the reported prosthesis wear or loosening cannot be confirmed from the reported information and the devices were not available for evaluation.This follow-up report is being submitted to relay corrected information.This follow-up report is being submitted to relay additional information.
|
| |
|