| Model Number |
GC-122 |
| Medical Device Problem Codes |
Break (1069); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y) gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36, (lot #d3d2368y); medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, preoperatively for a veterinarian, the thread was torn in the middle, and the needle with the needle holder was taken out of the packaging and then torn directly.A new device was used to resolve the issue.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This event has been reassessed and found not to be a reportable event and is not associated with a serious injury or potential for serious injury with reoccurrence.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h2, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.Thirty sealed devices were available for evaluation.The b reathable and foil pouches were inspected with light assisted microscopy with no abnormalities observed.Based on the functional testing performed on the representative samples, the condition of suture broken was detected.The most likely cause was determined to be manufacturing related.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|