• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD CAPROSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD CAPROSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC Back to Search Results
Model Number GC-122
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, preoperatively for a veterinarian, the thread was torn in the middle, and the needle with the needle holder was taken out of the packaging and then torn directly.A new device was used to resolve the issue.There was no patient involvement.
 
Additional Manufacturer Narrative
D10 concomitant product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); product: gc-122, gc-122 caprosyn* 3-0 und 75cm v20 x36 (lot#d3d2368y); medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: d9 (latest return date), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.Thirty sealed devices were available for evaluation.Visual inspection noted that the breathable and foil pouches were inspected with light assisted microscopy with no abnormalities observed.Functional packaging, moisture, visual and functional of suture was performed to assess potential suture degradation in package leading to the suture broken failure mode.It was reported that the suture had broken.The reported issue was confirmed.The most likely cause was determined to be manufacturing related.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPROSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18652123
Report Number9612501-2024-00334
Device Sequence Number12789151
Product Code GAM
UDI-Device Identifier10884521034426
UDI-Public10884521034426
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K032586
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model NumberGC-122
Device Catalogue NumberGC-122
Device Lot NumberD3D2368Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2024
Supplement Date Received by Manufacturer02/06/2024
11/12/2024
Initial Report FDA Received Date02/06/2024
Supplement Report FDA Received Date03/01/2024
11/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE NOTE ON H10.
Patient SexUnknown
-
-