| Catalog Number |
444150 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while using using bd synapsys¿ there was an application error that caused changes to the blood culture protocol time.This is a report of one occurrence, and no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.E1: initial reporter phone (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: this statement is to summarize the investigation of a complaint involving synapsys.It was reported synapsys was having issues with blood culture protocol times changing.No other issues were reported.Bd investigate the issue.Rule 8 should change the protocol time to 10 days, but the protocol time was changed back to 5 days.The lab was able to pick up this problem and put them back again on a new 5-day protocol.The customer was recommended to disable this rule and changed protocol times directly on fx.The instrument adapter was crashing from the system event logs.The dataupdate service was not sending anything to the instruments.The instrument adapter service was not crashing after 27th of february.The fx versions was updated.It was found that the fx versions were found to be running an older version.Capa will address this issue.This issue is caused by a software bug and will be fixed in a future synapsys release.This can be tracked by a system change request.This is a confirmed failure of a bd product.Review found complaints of this type were below statistical control for the month of january.Device history record review was not applicable as this is a standalone software product and the operation/functionality of this type of product is confirmed at the time of installation.Reports of this type will continue to be monitored.
|
| |
|
Event or Problem Description
|
|
It was reported that while using using bd synapsys¿ there was an application error that caused changes to the blood culture protocol time.This is a report of one occurrence, and no patient impact.
|
| |
|