Additional information: g3, h3, h6.Based on the information provided, which may include the device (if available and returned), pictures, videos, event descriptions, and any additional information from the field, arthrex concluded that the most likely cause of the reported failure is user error, specifically using the instrument to enlarge the surgical site.Directions for use dfu-0242 apollo probe.E.Precautions.5.Do not use excessive force when inserting or removing the rf probe.Do not insert the rf probe into an obstructed passageway.Patient injury and or product damage may occur.6.Do not use an rf probe as a lever to enlarge the surgical site or gain access to tissue, which may result in a bent or damaged rf probe.The complaint allegation was not confirmed as the device was not received for evaluation, and no evidence of the failure was provided.
|