• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH ALINITY C CREATININE REAGENT KIT; ALKALINE PICRATE, COLORIMETRY, CREATININE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH ALINITY C CREATININE REAGENT KIT; ALKALINE PICRATE, COLORIMETRY, CREATININE Back to Search Results
Catalog Number 07P99-20
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer reported falsely elevated alinity c creatinine and provided the following data: customer reference range is 50-98 umol/l.Sample id (b)(6) initial result was 1323.10 umol/l.Repeat result was 59.69 umol/l.Repeated result on another analyzer (b)(6) was 58 umol/l.The customer reported that all control levels for creatinine are in range.Per the customer the discrepant results were not reported out to the patient¿s medical provider.There was no reported impact to patient management.
 
Event or Problem Description
The customer reported falsely elevated alinity c creatinine and provided the following data: customer reference range is 50-98 umol/l.Sample id (b)(6)initial result was 1323.10 umol/l.Repeat result was 59.69 umol/l.Repeated result on another analyzer ((b)(6)) was 58 umol/l.The customer reported that all control levels for creatinine are in range.Per the customer the discrepant results were not reported out to the patient¿s medical provider.There was no reported impact to patient management.
 
Additional Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.Device history review did not identify any non-conformances, or deviations with the complaint lot.Ticket search by lot did not identify an increase in complaint activity for the complaint lot.Ticket trending review has not identified any trends.Labeling was reviewed and was found to address the customer reported issue.Based on the available information, no systemic issue or deficiency was identified for the alinity c creatinine lot number 88191un22.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY C CREATININE REAGENT KIT
Common Device Name
ALKALINE PICRATE, COLORIMETRY, CREATININE
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18655317
Report Number3002809144-2024-00030
Device Sequence Number17955845
Product Code CGX
UDI-Device Identifier00380740135102
UDI-Public00380740135102
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K083809
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/14/2024
Device Catalogue Number07P99-20
Device Lot Number88191UN22
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2024
Supplement Date Received by Manufacturer03/13/2024
Initial Report FDA Received Date02/06/2024
Supplement Report FDA Received Date03/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/11/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY C PROCESSING MODU, 03R67-01, (B)(6); ALNTY C PROCESSING MODU, 03R67-01, (B)(6)
Patient SexUnknown
-
-