| Catalog Number |
M0036129300 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3 other text : the device is not available to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that during the distal cervical segment of the right internal carotid artery dissecting pseudoaneurysm embolization procedure the subject coil deployed within the microcatheter.No clinical consequences were reported to the patient due to this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to the automated manufacturing execution system (mes), there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.No, additional information provided by the customer.While there are a number of potential causes for the reported issue, because review of available information failed to identify a definitive cause and the device was not returned for analysis, an assignable cause of undeterminable was assigned to the as reported "main coil prematurely detached/separated during use".
|
| |
|
Event or Problem Description
|
|
It was reported that during the distal cervical segment of the right internal carotid artery dissecting pseudoaneurysm embolization procedure the subject coil deployed within the microcatheter.No clinical consequences were reported to the patient due to this event.No additional information is available.
|
| |
|