• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Lot Number 0030159408
Medical Device Problem Codes Entrapment of Device (1212); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the stent got entrapped in the guide extension.The stenosed target lesion was located in the mildly tortuous and moderately calcified artery.A 3.50 x 12mm synergy xd drug-eluting stent was advanced for treatment.However, during insertion of the device the stent got stuck in the guide extension.Upon trying to pull out the device, the struts of the stent were damaged.The procedure was completed via alternative method.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
E1 initial reporter phone: (b)(6).
 
Event or Problem Description
It was reported that the stent got entrapped in the guide extension.The stenosed target lesion was located in the mildly tortuous and moderately calcified artery.A 3.50 x 12mm synergy xd drug-eluting stent was advanced for treatment.However, during insertion of the device the stent got stuck in the guide extension.Upon trying to pull out the device, the struts of the stent were damaged.The procedure was completed via alternative method.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
E1 initial reporter phone: (b)(6).Device evaluated by mfr.Synergy xd mr us 3.50 x 12mm stent delivery system (sds); catheter was returned for analysis.Visual, tactile, microscopic and dimensional testing was performed on the device.A visual, tactile examination identified no issues with the hypotube shaft.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues.A microscopic examination identified no issues with the hypotube shaft.A microscopic examination of the outer and inner lumen and mid-shaft section found no issues.Examination of the crimped stent via scope found evidence of stent damage.The proximal edge of the crimped stent had its struts lifted and bunched.There were no signs of movement and the stent was set between the proximal and distal markerbands.A microscopic examination of the bumper tip showed no signs of distal tip damage.The crimped stent od (outer diameter) was measured and was within maximum crimped stent profile measurement.No other device issues were identified during returned product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18655503
Report Number2124215-2024-01599
Device Sequence Number7111617
Product Code NIQ
UDI-Device Identifier08714729981039
UDI-Public08714729981039
Combination Product (Y/N)Y
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/05/2024
Device Lot Number0030159408
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/09/2024
Supplement Date Received by Manufacturer03/06/2024
Initial Report FDA Received Date02/06/2024
Supplement Report FDA Received Date03/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-