MERIT MEDICAL MANUFACTURING IMPRESS®.BRAIDED CATHETER; INTRAVASCULAR, DIAGNOSTIC CATHETER Back to Search Results
Catalog Number 510038VER
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The account alleges that during an endo vascular aortic repair [evar] procedure, a catheter tip detached in the patient's descending abdominal aorta.Excessive force was not used during catheter manipulations.The foreign body was successfully externalized via balloon capture technique, liberating the vessel along with all residual interventional equipment.No additional patient consequences to report.
 
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A supplemental report will be submitted once the evaluation is complete.
 
Additional Manufacturer Narrative
The suspect medical device has been returned for evaluation.The device was visually and microscopically investigated.The complaint is confirmed.The root cause could not be determined however, it is likely that significant force was applied to the device during clinical use.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®.BRAIDED CATHETER
Common Device Name
INTRAVASCULAR, DIAGNOSTIC CATHETER
Manufacturer (Section D)
MERIT MEDICAL MANUFACTURING
14646 kirby dr,
pearland TX 77047
Manufacturer (Section G)
MERIT MEDICAL MANUFACTURING
14646 kirby dr,
pearland TX 77047
Manufacturer Contact
bryson heaton bsn.rn
1600 merit pkwy
south jordan, UT 84095
8012531600
MDR Report Key18655608
Report Number3010665433-2024-00003
Device Sequence Number15480639
Product Code DQO
UDI-Device Identifier00884450009222
UDI-Public884450009222
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K053171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/11/2024
Device Catalogue Number510038VER
Device Lot NumberE2058224
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2024
Supplement Date Received by Manufacturer02/28/2024
Initial Report FDA Received Date02/06/2024
Supplement Report FDA Received Date03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INTERVENTIONAL GUIDEWIRE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-