| Catalog Number |
510038VER |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The account alleges that during an endo vascular aortic repair [evar] procedure, a catheter tip detached in the patient's descending abdominal aorta.Excessive force was not used during catheter manipulations.The foreign body was successfully externalized via balloon capture technique, liberating the vessel along with all residual interventional equipment.No additional patient consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A supplemental report will be submitted once the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect medical device has been returned for evaluation.The device was visually and microscopically investigated.The complaint is confirmed.The root cause could not be determined however, it is likely that significant force was applied to the device during clinical use.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
|
| |
|