• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number S-55-150-120-P6
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a chronic total occlusion in the right distal superficial femoral artery with moderate calcification.During preparation of the 5.5x150 mm supera self expanding stent system (sess), the device was loaded onto a guide wire which was inside the anatomy.The guide wire wire had lots of contrast and was sticky, causing resistance during advancement and removal.When the stent was pulled off the wire, the white tip separated.Therefore, the stent was not used in the patient and another same size supera was used to complete the procedure.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported material separation tip was able to be confirmed.The reported difficult to advance and the reported difficult to remove were unable to be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that a coagulation of blood and contrast on the guide wire resulted in the reported difficult to advance and the reported difficult to remove.Manipulation of the compromised device resulted in the noted multiple distal sheath kinks and ultimately resulted in the reported tip separation/noted inner member/ tip jacket separations; thus resulting in the noted partially deployed/flowered stent.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18655622
Report Number2024168-2024-01498
Device Sequence Number8245260
Product Code NIP
UDI-Device Identifier08717648226168
UDI-Public08717648226168
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Catalogue NumberS-55-150-120-P6
Device Lot Number3051061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/17/2024
Supplement Date Received by Manufacturer02/23/2024
Initial Report FDA Received Date02/06/2024
Supplement Report FDA Received Date03/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-