Catalog Number M0036126200
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an endovascular procedure, when the subject coil was deployed from the microcatheter and was being repositioned several times, the subject coil was prematurely detached from the delivery wire without using inzone.The coil was removed with the entire microcatheter and replaced with a new one the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
H3 other text : the device is not available to the manufacturer.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added.D4 expiration date - added.Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual testing as well as functional testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer indicated that the device was confirmed to be in good condition during preparation/prior to use on the patient and was prepared for use as per the dfu, continuous flush was set up and maintained throughout the clinical procedure, the coil was advanced slowly and gently, no torque was given where advancing the delivery wire, the coil was removed with the entire microcatheter, and the procedure was completed successfully using the same microcatheter and a new coil from the same lot.While there are a number of potential causes for the reported issue, because review of available information failed to identify a definitive cause and the device was not returned for analysis, an assignable cause of undeterminable was assigned to this complaint.
 
Event or Problem Description
It was reported that during an endovascular procedure, when the subject coil was deployed from the microcatheter and was being repositioned several times, the subject coil was prematurely detached from the delivery wire without using inzone.The coil was removed with the entire microcatheter and replaced with a new one the procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 6MM X 20CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key18655981
Report Number3008881809-2024-00047
Device Sequence Number17971117
Product Code HCG
UDI-Device Identifier07613327004625
UDI-Public07613327004625
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/02/2026
Device Catalogue NumberM0036126200
Device Lot Number24274475
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/17/2024
Supplement Date Received by Manufacturer02/07/2024
Initial Report FDA Received Date02/06/2024
Supplement Report FDA Received Date02/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN MICROCATHETER
Patient SexUnknown
-
-