Model Number CCWTT3
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Vitreous Floaters (1866); Blurred Vision (2137); Halo (2227)
Date of Event 12/05/2023
Type of Reportable Event Serious Injury
Event or Problem Description
A physician reported that following an intraocular lens (iol) implant procedure, the patients distance visual acuity was uncomfortable.Patient experienced hazy and halos.The near vision was good.The iol was perfectly centered and aligned.The surgeon observed some fine cortical wisps and wrinkling taken to operational room to clean.The fovea was perfect.The postop shows no improvement.The haptics was over rotated at 175 degrees.The surgeon asked patient to exchange for monofocal lens.Additional information has been requested.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was not returned.The complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.The product investigation could not identify a root cause for the reported complaint.The file was opened from a request made through the eomd program.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Follow up attempts were made for more details of the event.No further information has been provided at this time.File will be reopened when new information or the product is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TORIC TRIFOCAL UV ABSORBING IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18657916
Report Number1119421-2024-00211
Device Sequence Number8002286
Product Code MJP
UDI-Device Identifier00380652457439
UDI-Public00380652457439
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCCWTT3
Device Catalogue NumberCCWTT3.225
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/15/2024
Supplement Date Received by Manufacturer03/21/2024
Initial Report FDA Received Date02/07/2024
Supplement Report FDA Received Date04/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
-
-