| Model Number |
CCWTT3 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Codes |
Vitreous Floaters (1866); Blurred Vision (2137); Halo (2227)
|
| Date of Event |
12/05/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A physician reported that following an intraocular lens (iol) implant procedure, the patients distance visual acuity was uncomfortable.Patient experienced hazy and halos.The near vision was good.The iol was perfectly centered and aligned.The surgeon observed some fine cortical wisps and wrinkling taken to operational room to clean.The fovea was perfect.The postop shows no improvement.The haptics was over rotated at 175 degrees.The surgeon asked patient to exchange for monofocal lens.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned.The complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.The product investigation could not identify a root cause for the reported complaint.The file was opened from a request made through the eomd program.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Follow up attempts were made for more details of the event.No further information has been provided at this time.File will be reopened when new information or the product is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|