| Catalog Number |
ASK-04001-VCUH3 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "nurse was attempting to advance the peel-away into a young patient with tough skin and the peel-away accordioned about 1 cm below the proximal end.The sheath was replaced with a gault sheath and the procedure was completed without incident".No patient harm or injury.The patient's current condition reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The actual device was not returned; however, the customer provided one photo for analysis.The photo shows a peel-away sheath/dilator assembly with signs of crimping on the peel-away sheath body and tip.Therefore, the complaint of peel-away sheath body damaged was able to be confirmed by the photo, however, a complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "do not use excessive force when introducing guidewire, peel-away sheath over tissue dilator, or tissue dilator as this can lead to vessel perforation and bleeding." without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported "nurse was attempting to advance the peel-away into a young patient with tough skin and the peel-away accordioned about 1 cm below the proximal end.The sheath was replaced with a gault sheath and the procedure was completed without incident".No patient harm or injury.The patient's current condition reported as "fine.".
|
| |
|