The event is reported at this time based on analysis results which indicate a possibly reportable event.H3.Product analysis: equipment used: vis (m-78210), 203cm ruler (m-83361) as found condition (condition of returned device): the concerto implant coil was returned for analysis within a shipping box; within a sealed biohazard pouch; within an opened concerto implant coil outer carton and within an opened concerto implant coil inner pouch.The concerto implant coil was returned without introducer sheath.Visual inspection/damage location details: the concerto coil pushwire was found to be broken but retained by release wire at break indicator and bent at ~69.5cm from proximal end.The implant coil was found to be detached and not returned.No other anomalies were observed.Conclusion: based on the device analysis and the reported information, the customer¿s report of ¿coil resistance/stuck in sheath¿ was unable to be confirmed as the concerto was returned with implant coil already detached and without introducer sheath.There were no reported issues pushing the concerto coil out from within its introducer sheath during preparation.The root cause could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|