• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION POLARIS ULTRA; STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION POLARIS ULTRA; STENT, URETERAL Back to Search Results
Model Number M0061921420
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0406 captures the reportable event of stent kinked, inside the patient.
 
Event or Problem Description
It was reported to boston scientific corporation that a polaris ultra ureteral stent was used to treat kidney stones during a percutaneous nephrolithotomy procedure in the left ureter performed on (b)(6) 2024.During insertion, it was noted that at the end of the procedure, when the stent was inserted and pushed, it was found kinked.The procedure was successfully completed with another polaris ultra ureteral stent.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that a polaris ultra ureteral stent was used to treat kidney stones during a percutaneous nephrolithotomy procedure in the left ureter performed on (b)(6) 2024.During insertion, it was noted that at the end of the procedure, when the stent was inserted and pushed, it was found kinked.The procedure was successfully completed with another polaris ultra ureteral stent.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0406 captures the reportable event of stent kinked, inside the patient.Update to block d4: expire date 08/09/2026.Update to block g2: costa rica.Block h10: the returned polaris ultra ureteral stent was analyzed, and a visual evaluation noted that the bladder tip was buckled.During magnification, it was observed closely that the bladder coil was buckled, tip was torn, and suture hole was torn.No other problems with the device were noted.The reported event was not confirmed.Taking all available information into consideration, most likely, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.It is possible to conclude that this problem could be caused by operational factors.Interaction of the device between the positioner and the guide wire while the device was pushing up, such as excess of force applied over the device during the procedure and could have contributed to the failure.For the reported problem of stent kinked, was not detected during the analysis.For this reason, the as analyze code will be no problem detected.Therefore, all compiled information on this investigation determines that the most probable cause is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARIS ULTRA
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18661598
Report Number2124215-2024-06180
Device Sequence Number12544871
Product Code FAD
UDI-Device Identifier08714729124542
UDI-Public08714729124542
Combination Product (Y/N)N
Initial Reporter CountryCS
PMA/510(K) Number
K010002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/09/2026
Device Model NumberM0061921420
Device Catalogue Number192-142
Device Lot Number0032188018
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/18/2024
Supplement Date Received by Manufacturer03/12/2024
Initial Report FDA Received Date02/07/2024
Supplement Report FDA Received Date04/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age38 YR
Patient SexMale
-
-