Catalog Number 1012625-29
Medical Device Problem Codes Leak/Splash (1354); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.It was reported that the omnilink 9.0x29 was delivered to the lesion and air aspiration was performed.It should be noted that the omnilink elite instructions for use (ifu), prepare an inflation device / syringe with diluted contrast medium.Attach an inflation device / syringe to the stopcock; attach to the inflation port.With the tip down, orient the delivery system vertically.Open the stopcock to delivery system; pull negative for 30 seconds; release to neutral for contrast fill.Close the stopcock to the delivery system; purge the inflation device / syringe of all air.Repeat steps until all air is expelled.In this case, it is unknown the ifu deviation contributed to the reported event.A conclusive cause for the reported leak/splash cannot be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling; therefore, no product-related corrective action will be implemented in this case.H6: 2017 device code clarifier- failure to follow steps / instructions.
 
Event or Problem Description
It was reported the procedure was to treat the subclavian artery.The 9.0x29mm otw omnilink elite stent delivery system (sds) was advanced to the lesion and air aspiration performed; however more air than usual was noted therefore a leak was suspected.The sds was removed without issue and the procedure completed using another omnilink.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18666988
Report Number2024168-2024-01588
Device Sequence Number8331945
Product Code NIO
UDI-Device Identifier08717648179334
UDI-Public08717648179334
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Catalogue Number1012625-29
Device Lot Number3020241
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/22/2024
Initial Report FDA Received Date02/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-