| Model Number |
7840 |
| Medical Device Problem Codes |
Difficult to Advance (2920); Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that foreign material was present in the device.An expo diagnostic catheter was selected for use.During the procedure, the catheter was advanced in patient.However, an attempt to load the wire, the wire was not successful due to a white plastic piece being lodged in the catheter.The procedure was completed using an alternate device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that foreign material was present in the device.An expo diagnostic catheter was selected for use.During the procedure, the catheter was advanced in patient.However, an attempt to load the wire, the wire was not successful due to a white plastic piece being lodged in the catheter.The procedure was completed using an alternate device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was returned for analysis.Visual and microscopic inspection found inner liner delamination.A photo provided by the customer depicted the catheter with the tip cut and the inner liner separated next to the tip.
|
| |
|