Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "the liquid does not come out of the syringe" there was no patient harm or consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The reported complaint of "occlusion - blockage/part stuck" was confirmed based upon the sample received.The customer returned one-unit mad100 mad nasal with 3 ml syringe for investigation.The returned sample was visually examined with and without magnification.Visual examination of the returned device revealed that the sample appears typical.The syringe was returned with the mad device.A functional inspection was performed on the returned device.The returned syringe was filled with water to test the atomization of the returned device.Upon testing, the device was not able to atomize properly.The water came out as a continuous stream.The device was sent to the manufacturing site for further analysis.The manufacturing site inspected the device and found no issues with the positioning of the insert (component that helps with atomization).It could not be determined what prevented the device from atomizing properly at this time.A device history record review was performed with no evidence to suggest a manufacturing related cause.Teleflex will continue to monitor and trend on this issue.
|
| |
|