• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL MAD NASAL WITH 3 ML SYRINGE; NEBULIZER, MEDICINAL, NON-VENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL MAD NASAL WITH 3 ML SYRINGE; NEBULIZER, MEDICINAL, NON-VENT Back to Search Results
Model Number IPN048826
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "the liquid does not come out of the syringe" there was no patient harm or consequence.
 
Additional Manufacturer Narrative
(b)(4).The reported complaint of "occlusion - blockage/part stuck" was confirmed based upon the sample received.The customer returned one-unit mad100 mad nasal with 3 ml syringe for investigation.The returned sample was visually examined with and without magnification.Visual examination of the returned device revealed that the sample appears typical.The syringe was returned with the mad device.A functional inspection was performed on the returned device.The returned syringe was filled with water to test the atomization of the returned device.Upon testing, the device was not able to atomize properly.The water came out as a continuous stream.The device was sent to the manufacturing site for further analysis.The manufacturing site inspected the device and found no issues with the positioning of the insert (component that helps with atomization).It could not be determined what prevented the device from atomizing properly at this time.A device history record review was performed with no evidence to suggest a manufacturing related cause.Teleflex will continue to monitor and trend on this issue.
 
Event or Problem Description
It was reported that "the liquid does not come out of the syringe" there was no patient harm or consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAD NASAL WITH 3 ML SYRINGE
Common Device Name
NEBULIZER, MEDICINAL, NON-VENT
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18668185
Report Number3003898360-2024-00213
Device Sequence Number15543228
Product Code CCQ
UDI-Device Identifier14026704613103
UDI-Public14026704613103
Combination Product (Y/N)N
Initial Reporter CountryIT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/22/2025
Device Model NumberIPN048826
Device Catalogue NumberMAD100
Device Lot Number73H2200671
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2024
Supplement Date Received by Manufacturer03/19/2024
Initial Report FDA Received Date02/08/2024
Supplement Report FDA Received Date03/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-