| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the instruments were discovered fractured and missing a bearing during the instrument check-in process.Attempts have been made and all available information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2024-00440; 0001822565-2024-00441; 0001822565-2024-00442; 0001822565-2024-00443; 0001822565-2024-00445; 0001822565-2024-00446.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6.H6: tasp bottom- mechanical (g04) - provisional bottom.Performed a visual inspection of the returned item and found it to exhibit signs of repeated use nicked / gouged / wear and has fractured on the medial side of the post feature.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.The root cause of the reported issue is attributed to the over 6+ years in the field.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Upon review of our reporting decision on fracture of articular surface provisionals.A review of all complaints for this product indicates that there has never been an event that has resulted in a death or serious injury.Therefore, zimmer biomet will not be reporting this failure mode going forward.
|
| |
|