• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12°

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12° Back to Search Results
Model Number WA22302D
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2023
Event Type  malfunction  
Event Description
It was observed during the device evaluation that the tip of the electrode was fused.The event was found during reprocessing.There were no reports of patient involvement.This medical device report (mdr) is being submitted to capture the reportable malfunction found during the evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation, and the evaluation found that there was a dent in the blue part of electrode on the opposite side, not in the tip.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
 
Manufacturer Narrative
Correction: h8 was inadvertently omitted from the initial report.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to thermal stress from usage-related wear and tear.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12°
Type of Device
HF-RESECTION ELECTRODE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg/de/germany 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg/de/germany
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18670852
MDR Text Key334921303
Report Number9610773-2024-00398
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761051665
UDI-Public14042761051665
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA22302D
Device Lot Number1000108623
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/05/2024
Initial Date FDA Received02/08/2024
Supplement Dates Manufacturer Received05/03/2024
Supplement Dates FDA Received05/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-