| Catalog Number |
305945 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the bd safetyglide safety mechanism broke.The following information was provided by the initial reporter: "the entire safety device slide up and off the syringe when the nurse deployed it.I wanted to bring it to your attention in case there has been any other incidents.Staff may not report it if it occurs when there is no negative consequence.The rn discarded it but showed the charge nurse prior to doing so.Charge nurse verified her account.".
|
| |
|
Additional Manufacturer Narrative
|
|
H.3: if a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: one photo was received verifying the customer's complaint of safety mechanism breaking off.The photo clearly presents the issue at hand but without further information like a physical sample for further investigation, the root cause remains unknown.The production history was analyzed and there were no issues found during quality inspections that would lead to this issue.
|
| |
|
Event or Problem Description
|
|
Photo received.
|
| |
|
Search Alerts/Recalls
|