MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977762
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The used company cartridge was returned.The lens was not returned.The cartridge was microscopically examined.Inadequate viscoelastic was observed in the cartridge.A small amount of viscoelastic was only observed just past the loading area.No viscoelastic was observed in the nozzle or the cartridge tip.No damage or abnormalities were observed.The cartridge had evidence of placement into a handpiece.The used company cartridge was cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.Product history records were reviewed and documentation indicated the product met release criteria.A qualified lens model/diopter, viscoelastic and handpiece were indicated.Root cause of the reported lens damage could not be verified.The lens was not returned.The root cause for the reported damage may be related to a failure to follow the ifu (instructions for use).Inadequate viscoelastic was observed in the cartridge.The returned used company cartridge was evaluated and no damage or abnormalities were observed.Top coat dye stain testing was conducted with acceptable results.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd (ophthalmic viscosurgical device) in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage the manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-health care professional reported that during an intraocular lens (iol) implant procedure, reported with a description of observed scratches on a lens when surgeon implanted lens.The customer believes it is due to the cartridge that was used.The surgeon mentioned that this has happened on more than one occasion.Additional information is received.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C
Common Device Name
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18674676
Report Number1119421-2024-00223
Device Sequence Number7395885
Product Code MSS
UDI-Device Identifier00380659777622
UDI-Public00380659777622
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K112977
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/03/2026
Device Catalogue Number8065977762
Device Lot Number15719334
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/17/2024
Initial Report FDA Received Date02/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/05/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACRYSOF SINGLEPIECE IOL; MCPHERSON FORCEPS; MONARCH III INJECTOR; PROVISC
Patient SexUnknown
-
-