| Model Number |
11327600 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Siemens became aware of a malfunction that occurred while operating the artis icono biplane system.During an emergency procedure, the system repeatedly went into bypass mode.The patient was moved to a different room and the procedure was completed on an alternate unit.We have no indications of any adverse effects on the health status of the involved patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens is conducting a thorough investigation of the reported event.A supplement report will be filed if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the reported issue was completed by our experts.The onsite service inspection identified an issue with the large display.The large display was exchanged as part of service activity, which resolved the problem.Detailed investigation of the defective part revealed that the large display's backlight sensor module was defective, resulting in a decrease in luminance after 15 minutes of operation.Eventually, this caused the system to go into "bypass" mode.The occurrence rate of the aforementioned error pattern was checked.A possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.
|
| |
|