| Catalog Number |
0167L24 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
01/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had 3-way catheter placed in the operating room and the patient came to the post anesthesia care unit and the foley catheter balloon had ruptured and the catheter in urethra per md.Catheter needed to be replaced and the foley clotted off.Doctor needed to do irrigation and the foley changed and the tubing from continuous bladder irrigation not flowing properly and this also changed out and patient was having a open retropubic simple prostatectomy and pain was reported in the bladder.No medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.However, the potential root cause for this failure mode could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/ operator error.A potential root cause for this failure mode could be due to thin sac that may lead to burst balloon.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: "bard® ez-lok® sampling port (indicated by the blue stem in the port) accepts luer lock (fig.1a) or slip tip syringes (fig.1b).1.Occlude drainage tubing a minimum of 3 inches below the sampling port by kinking the tubing until urine is visible under the access site.2.Swab surface of bard® ez-lok® sampling port with antiseptic wipe.(fig.2) 3.Using aseptic technique, position the syringe in the center of the sampling port.The syringe should be held perpendicular to the surface of the sampling port (at approximately 80 - 100 degree angle).Press the syringe firmly and twist gently to access the sampling port.(fig.3) 4.Slowly aspirate urine sample into syringe and remove syringe from sample port.(fig.4) 5.Unkink tubing and transfer urine specimen into specimen cup or follow hospital protocol.Discard syringe according to hospital protocol.6.Follow established hospital protocol for specimen labeling and transport to lab.Sterile: contents of inner wrap are sterile unless package has been opened or damaged.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not re-sterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had 3-way catheter placed in the operating room and the patient came to the post anesthesia care unit and the foley catheter balloon had ruptured and the catheter in urethra per md.Catheter needed to be replaced and the foley clotted off.Doctor needed to do irrigation and the foley changed and the tubing from continuous bladder irrigation not flowing properly and this also changed out and patient was having a open retropubic simple prostatectomy and pain was reported in the bladder.No medical intervention was reported.Per additional information received on 26feb2024, they believed on further investigation it was the drain bag might have been blocked not allowing gravity to work.The patient was taken back for a follow up procedure to remove blood clots.
|
| |
|