| Model Number |
133632 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
At this time, the device is expected to be returned to orthalign for investaion, but has not been recieved.If the device is recieved, an investigation will be performed to confirm the problem and determine the root cause, and a follow-up report will be filed detailing the conclusions.This report is being filed out of an abundane of caution and with an understanding of the patient harm that could be caused by device inaccuracy.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon felt the v/v angle on the femur was not accurate.The knee turned out well and there was no impact to patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon felt the v/v angle on the femur was not accurate.The knee turned out well and there was no impact to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The navigation unit and the reference sensor were returned to orthalign for evaluation and the investigation has been completed by an orthalign engineer.The behavior was not reproducable by the orthalign engineer and the event was unable to be confirmed.The navigation unit and reference sensor passed all testing and no device defect was found.The reported issue and possible explanations are addressed in orthalign's risk documentation.Orthalign will continue to monitor the complaint data and take action if necessary.No further action is required at this time.
|
| |
|