| Model Number |
8900-0225-01 |
| Medical Device Problem Codes |
Failure to Deliver Shock/Stimulation (1133); Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the associated device failed to discharge using these electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the associated device failed self-test using these electrode pads.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section b5 and h6 (device problem code).The customer's report was verified and attributed to a disconnected jumper wire (self test wire) on the electrode pad.This type of failure does not disable the electrode from delivering therapy when attached to a defibrillator.This is a malfunction which only impacts self-test of the electrode with the defibrillator.This type of malfunction does not pose any clinical impact and does not meet our guidelines for reportability.
|
| |
|