• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-DETEK INCORPORATED ELECTRODES, ONESTEP CPR AA, SINGLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-DETEK INCORPORATED ELECTRODES, ONESTEP CPR AA, SINGLE Back to Search Results
Model Number 8900-0225-01
Medical Device Problem Codes Failure to Deliver Shock/Stimulation (1133); Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that during biomed testing, the associated device failed to discharge using these electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Event or Problem Description
Complainant alleged that during biomed testing, the associated device failed self-test using these electrode pads.
 
Additional Manufacturer Narrative
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section b5 and h6 (device problem code).The customer's report was verified and attributed to a disconnected jumper wire (self test wire) on the electrode pad.This type of failure does not disable the electrode from delivering therapy when attached to a defibrillator.This is a malfunction which only impacts self-test of the electrode with the defibrillator.This type of malfunction does not pose any clinical impact and does not meet our guidelines for reportability.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRODES, ONESTEP CPR AA, SINGLE
Common Device Name
ELECTRODE
Manufacturer (Section D)
BIO-DETEK INCORPORATED
525 narragansett park drive
pawtucket RI 02861
Manufacturer Contact
tina lombari
525 narragansett park drive
pawtucket, RI 02861
4017291400
MDR Report Key18677116
Report Number1218058-2024-00012
Device Sequence Number6195538
Product Code MKJ
UDI-Device Identifier00847946016289
UDI-Public00847946016289
Combination Product (Y/N)N
Initial Reporter CountryKS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/25/2025
Device Model Number8900-0225-01
Device Catalogue Number8900-0225-01
Device Lot Number2623B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2024
Supplement Date Received by Manufacturer01/19/2024
Initial Report FDA Received Date02/09/2024
Supplement Report FDA Received Date03/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-