ZOLL MEDICAL CORPORATION PROPAQ MD DEFIBRILLATOR Back to Search Results
Model Number PROPAQ MD
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Complainant alleged that during biomed testing, the device's screen turned completely white and was illegible.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Additional Manufacturer Narrative
The customer's report was duplicated and attributed to a faulty lcd display.The lcd module was replaced to resolve the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPAQ MD DEFIBRILLATOR
Common Device Name
PROPAQ MD
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key18677240
Report Number1220908-2024-00415
Device Sequence Number20637201
Product Code MKJ
UDI-Device Identifier00847946020156
UDI-Public00847946020156
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P160022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberPROPAQ MD
Device Catalogue NumberPROPAQ MD
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2024
Supplement Date Received by Manufacturer01/19/2024
Initial Report FDA Received Date02/09/2024
Supplement Report FDA Received Date03/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-