| Catalog Number |
MS25643 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that on (b)(6) 2024 at approximately 9am, the draeger m540 will not stay connected to the c700 cockpit.It was noticed that there was communication lost to c700 from m540, and the m540 was replaced with a backup unit and the case continued with no further issues.There was no adverse event or patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A follow-up report will be submitted upon completion of this investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported that on (b)(6) 2024 at approximately 9am, the draeger m540 will not stay connected to the c700 cockpit.It was noticed that there was communication lost to c700 from m540, and the m540 was replaced with a backup unit and the case continued with no further issues.There was no adverse event or patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported to draeger that the m540 disconnected from the cockpit.The m540 was returned to drager for repair, and it was determined that the mainboard was not recognizing the wireless card, causing the loss of communication.Root cause was identified as faulty mainboard.The mainboard was replaced to resolve the issue.The m540 was tested and returned to the customer.No further issues have been reported.
|
| |
|