• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRAEGER MEDICAL SYSTEMS, INC INFINITY ACUTE CARE SYSTEM (M540); PHYSIOLOGIC MONITORING SYSTEM, SINGLE-PATIENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRAEGER MEDICAL SYSTEMS, INC INFINITY ACUTE CARE SYSTEM (M540); PHYSIOLOGIC MONITORING SYSTEM, SINGLE-PATIENT Back to Search Results
Catalog Number MS25643
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2024
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that on (b)(6) 2024 at approximately 9am, the draeger m540 will not stay connected to the c700 cockpit.It was noticed that there was communication lost to c700 from m540, and the m540 was replaced with a backup unit and the case continued with no further issues.There was no adverse event or patient injury reported.
 
Additional Manufacturer Narrative
A follow-up report will be submitted upon completion of this investigation.
 
Event or Problem Description
The customer reported that on (b)(6) 2024 at approximately 9am, the draeger m540 will not stay connected to the c700 cockpit.It was noticed that there was communication lost to c700 from m540, and the m540 was replaced with a backup unit and the case continued with no further issues.There was no adverse event or patient injury reported.
 
Additional Manufacturer Narrative
It was reported to draeger that the m540 disconnected from the cockpit.The m540 was returned to drager for repair, and it was determined that the mainboard was not recognizing the wireless card, causing the loss of communication.Root cause was identified as faulty mainboard.The mainboard was replaced to resolve the issue.The m540 was tested and returned to the customer.No further issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY ACUTE CARE SYSTEM (M540)
Common Device Name
PHYSIOLOGIC MONITORING SYSTEM, SINGLE-PATIENT
Manufacturer (Section D)
DRAEGER MEDICAL SYSTEMS, INC
3135 quarry road
telford PA
Manufacturer (Section G)
DRAEGER MEDICAL SYSTEMS, INC
3135 quarry road
telford PA
Manufacturer Contact
3135 quarry road
telford, PA 
2156602400
MDR Report Key18677304
Report Number1220063-2024-00013
Device Sequence Number12611266
Product Code MHX
UDI-Device Identifier04049098054409
UDI-Public(01)04049098054409(11)141219(93)MS25643-05
Combination Product (Y/N)N
Initial Reporter StateTX
PMA/510(K) Number
K113798
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 05/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMS25643
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/09/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/12/2024
Supplement Date Received by Manufacturer05/10/2024
Initial Report FDA Received Date02/09/2024
Supplement Report FDA Received Date05/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NA.
Patient SexUnknown
-
-