| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/04/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient was inadvertently exposed to the treatment fraction for 5.2 seconds when imaging was intended.
|
| |
|
Additional Manufacturer Narrative
|
|
The treatment tab was opened before the mvct (mega volt computed tomography) was acquired.The user thought he was viewing the scan screen, so with the treat screen displayed, he turned the key to image and then pressed the start button on the status console.The patient was exposed to the treatment fraction for 5.2 seconds when imaging was intended.It is not yet known at this time if there wa a serious injury to the patient.Please note, this report was initially submitted under manufacturer number 3003893069-2023-00018 on 2023-dec-15.
|
| |
|
Additional Manufacturer Narrative
|
|
An analysis was completed of the dose deviation that the patient received.The result of that analysis indicated some deviations in the intended dose for delivery in the specific fraction, but negligible deviation from the intended prescription dose, both to the target and to the oars (organs at risk).The incident was assessed as user error, particularly because the user continued on the treat panel even though they were intending to scan the patient for positioning.The overall risk was acceptable and there was no serious injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi and date of manufacture correction: during a review it was discovered that the udi and date of manufacture was incorrectly typed into the service work order system.The information has now been corrected in the system and this report is being updated.See d4 and h4 of this report.
|
| |
|