• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LNR, NEUTRAL, 32MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LNR, NEUTRAL, 32MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number NV GXL LNR, NEUTRAL, 32MM ID, GROUP 3 CUPS
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section d10: concomitant products: crown cup inlay,lat.+5 gr.3,32mm (cat# 136-32-53 / serial# (b)(6)).Crown cup,cluster-hole gr.56 (cat# 180-01-56 / serial# (b)(6)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported by the germany competent authorities and as part of the manufacturer's recall campaign, the patient went to have both hip prostheses checked (left implanted in 2019, right in 2021).The x-ray and ct check showed clear decentration both prosthetic heads with unusually large osteolysis in both acetabula as a sign of premature inlay wear.First, the right side was revised on (b)(6) 2023.The suspected diagnosis was confirmed and the inlay was replaced with a specially approved vitd-hardened one inlay (novation xle, neutral liner, group 3, 36 mm i.D., ref 140-36-53, sn (b)(6)).As part of the exchange operation, in addition to the inlay exchange, the solid cup integrity was determined as well as curettage and sealing of the cysts in the socket using allogeneic cancellous bone and changing the prosthetic head.The explants are currently with hospital and will be returned to the patient at their own request handed out.No further information provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information: please disregard this report as it was submitted in error.Event was previously reported under report # 1038671-2024-00055.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, NEUTRAL, 32MM ID, GROUP 3 CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
miguel soza
MDR Report Key18679010
Report Number1038671-2024-00188
Device Sequence Number6155770
Product Code JDI
UDI-Device Identifier10885862022172
UDI-Public10885862022172
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 01/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/20/2022
Device Model NumberNV GXL LNR, NEUTRAL, 32MM ID, GROUP 3 CUPS
Device Catalogue Number130-32-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2024
Supplement Date Received by Manufacturer01/14/2026
Initial Report FDA Received Date02/09/2024
Supplement Report FDA Received Date01/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/21/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1729-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-