Model Number 2ACH20
Medical Device Problem Codes Signal Artifact/Noise (1036); Entrapment of Device (1212); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Physical Entrapment (2327)
Date of Event 02/06/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a cryo ablation procedure, the integrated dilator/needle was not able to lock into the sheath.The dilator/needle was replaced which resolved the issue.After the first ablation in the right superior pulmonary vein (rspv), the mapping catheter had a lot of noise.The mapping catheter cable replaced, and then the pin block was replaced without resolution.The physician attempted to reposition the mapping catheter and it was noted that it did not move at all.When the physician pulled the mapping catheter all the way back, the end of mapping catheter remained lodged in the right superior pulmonary vein (rspv).The balloon catheter was removed; a snare was inserted, and the end of the mapping catheter was snared, pulled back into the sheath, and removed.The case was completed with cryo.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product event summary: the mapping catheter (b)(6) of lot 8779916 and data files containing an image were returned and analyzed.The returned image showed a broken shaft and wires separated on two parts inside a bag.Visual inspection of the shaft segment area showed the shaft was broken approximately 57.5 inches from the lemo connector, inspection also identified broken wires at the broken shaft area.Functional testing was unable to be carried out due to the condition of the mapping catheter.In conclusion, the reported mapping catheter tip entrapment could not be confirmed through analysis.The mapping catheter failed the return product inspection due to a broken shaft and broken electrode wire in the shaft segment.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE ADVANCE MAPPING CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18680541
Report Number2182208-2024-00537
Device Sequence Number20667296
Product Code DRF
UDI-Device Identifier00643169590984
UDI-Public00643169590984
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K162892
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/03/2026
Device Model Number2ACH20
Device Catalogue Number2ACH20
Device Lot Number8779916
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2024
Supplement Date Received by Manufacturer02/13/2024
Initial Report FDA Received Date02/09/2024
Supplement Report FDA Received Date03/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-