| Model Number |
WA64700A |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
It was reported, the reusable needle holder jaws were missing.The issue occurred during reprocessing.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.A definitive root cause was not identified; however, based on the results of the investigation, the probable cause of the malfunction would likely be that the issue was most likely caused by excessive force, such as improper handling.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|