• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", 5 X 330 MM, NEEDLE HOLDER, STRAIGHT; REUSABLE NEEDLE HOLDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", 5 X 330 MM, NEEDLE HOLDER, STRAIGHT; REUSABLE NEEDLE HOLDER Back to Search Results
Model Number WA64700A
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for evaluation.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event or Problem Description
It was reported, the reusable needle holder jaws were missing.The issue occurred during reprocessing.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.A definitive root cause was not identified; however, based on the results of the investigation, the probable cause of the malfunction would likely be that the issue was most likely caused by excessive force, such as improper handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAWS "HIQ+", 5 X 330 MM, NEEDLE HOLDER, STRAIGHT
Common Device Name
REUSABLE NEEDLE HOLDER
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18681299
Report Number9610773-2024-00414
Device Sequence Number7217429
Product Code GCJ
UDI-Device Identifier04042761031097
UDI-Public04042761031097
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K923982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA64700A
Device Lot Number177W-0039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2024
Supplement Date Received by Manufacturer09/03/2024
Initial Report FDA Received Date02/09/2024
Supplement Report FDA Received Date09/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-